<exptProc>
<run>
<name>Introduction</name>
	<step>
	<caption>Introduction</caption>
	<video>./data/expts/ee001/Exp1_intro1.flv</video>
		<description><![CDATA[<div class =tdText><strong>Specification of Instruments Used</strong><ul><li>2 D.C Ammeters<ol type=i><li>0 &ndash; 2.5A/5A (Test)</li><li>0 &ndash;  5A/10A (Standard)</li></ol></li><li>Rheostat 100 &Omega; / 5A Rating</li><li>Load Box 0-20A in steps of 1A, 110V Rating</li><li>2 D.C Voltmeters<ol type=i><li>0 &ndash; 75V/150V (Test)</li><li>0 &ndash; 75V/150V (Test)</li></ol></li></ul></div></div>]]></description>
	</step>
</run>
<run>
<name>Ammeter Calibration</name>
	<step>
	<caption>Step 1</caption>
	<video>./data/expts/ee001/Exp1_R1_Stap1.flv</video>
		<description><![CDATA[<div class =tdText><ul><li>Make the connections using the board as per the schematic circuit diagram. </li><li> When the connections are complete, connect the power supply and then switch on the power.</li></ul></div>]]></description>
	</step>
	<step>
	<caption>Step 2</caption>
	<video>./data/expts/ee001/Exp1_R1_Stap2.flv</video>
		<description><![CDATA[<div class =tdText><ul><li>With no load applied, check whether the test ammeter has any zero error. </li><li> Adjust R<sub>1</sub> (LoadBox) and R<sub>2</sub> (Rheostat) to get the pointer of the test ammeter on desired markings on its scale. Note the corresponding reading in standard ammeter.</li></ul></div>]]></description>
	</step>
</run>

<run>
<name>Voltmeter Calibration</name>
<step>
<caption>Step 1</caption>
<video>./data/expts/ee001/Exp1_R2_Stap1.flv</video>
		<description><![CDATA[<div class =tdText><ul><li>Connect the meters as per circuit diagram. </li><li>Correct the voltmeters for zero errors, if necessary.</li></ul></div>]]></description>
</step>
<step>
<caption>Step 2</caption>
<video>./data/expts/ee001/Exp1_R2_Stap2.flv</video>
		<description><![CDATA[<div class =tdText><ul><li>Check &nbsp;whether the  voltmeter has any zero error. </li><li>Adjust R using slider of the rheostat to get the pointer of the test voltmeter on desired markings on its scale.<br>Note the corresponding readings in standard voltmeter. </li><li>Perform the same procedure until the full scale of the test voltmeter is reached.</li></ul></div>]]></description>

</step>
</run>

</exptProc>